| dc.identifier.citation | Raquel Vaquero-Cristóbal, Ana Catarina-Moreira, Francisco Esparza-Ros, Carlos Barrigas, Mario Albaladejo-Saura & Filomena Vieira (2023) Skinfolds compressibility and digital caliper’s time response in skinfold measurement in male and female young adults, Journal of the International Society of Sports Nutrition, 20:1, 2265888, DOI: 10.1080/15502783.2023.2265888 | es |
| dc.identifier.uri | http://hdl.handle.net/10952/8300 | |
| dc.description.abstract | Background: The skinfold caliper reading of the skinfold thickness
depends on its dynamic compressibility. This has led to the fact
that, while it is indicated that skinfold readings should be taken
when the reading is stable, there is no consensus on at what second
the reading should be taken after the application of the skinfold
caliper. The new Lipowise PRO digital skinfold caliper was used to
analyze the evolution of skinfold readings under skinfold caliper
pressure. The aim of the present investigation were: a) to analyze
the evolution of the reading time of individual skinfolds when
subjected to skinfold caliper pressure and when the skinfold reading
reaches stability; b) to describe the physical behavior of skinfold
tissues’ time response to skinfold caliper pressure, and to explore
differences between sites and subjects’ skinfolds compressibility;
and c) to analyze the sex differences in both the reading and the
evolution of the skinfold over time.
Methods: A descriptive cross-sectional design was followed with
a convenience sample of 165 healthy young adults (79 males and
86 females), with eight skinfolds measured using the Lipowise PRO
skinfold caliper. The Lipowise PRO skinfold caliper uses
a programmable reading time allowing for the measurement of
the skinfold’s thickness at a rate of 100 times per second, and
monitoring skinfold behavior over the 3-second measurement period,
thereby enabling the assessment of the tissue response to the
constant force exerted by the skinfold caliper jaws.
Results: All skinfolds showed statistical differences in terms of
compressibility characteristics (p < 0.001). Significant differences
were found between measurement time points for individual skinfolds
and sum of skinfolds (p < 0.001–0.025). Stabilization being
found depending on the skinfold measured from 1.5 seconds for
biceps, subscapular, iliac crest, supraspinale, abdominal, and thigh skinfolds; 2.0 seconds for Σ6 and Σ8 skinfolds; and 2.5 seconds for
triceps and calf skinfolds. It was observed an effect of sex on this
issue (p < 0.001–0.030). More specifically, in the case of males, the
supraspinale and abdominal skinfolds stabilized after 1.5 seconds;
the calf skinfold and Σ6 and Σ8 skinfolds stabilized after 2 seconds;
while the rest of the skinfolds did not stabilize until 3 seconds. In
the case of females, no stabilization of the triceps skinfold was
found, while the rest of the individual skinfolds and the Σ6 and Σ8
skinfolds stabilized from 1.5 seconds. A regression analysis indicated
that skinfold thickness could be predicted based on measurement
time in 50–77% of the cases (p = 0.001).
Conclusion: A skinfold caliper application, using the digital caliper
Lipowise PRO, of three seconds may be sufficient for achieving
stability in the measurement and for obtaining the minimum
value for most individual and sum of skinfolds. However, there are
certain skinfolds that may require more time when performed on
certain individuals, which vary according to sex. | es |
| dc.language.iso | en | es |
| dc.rights | Attribution-NonCommercial-NoDerivatives 4.0 Internacional | * |
| dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | * |
| dc.subject | Skinfold thickness | es |
| dc.subject | Anthropometry | es |
| dc.subject | Body composition | es |
| dc.subject | Response time | es |
| dc.subject | Skinfold caliper | es |
| dc.subject | Fat mass | es |
| dc.title | Skinfolds compressibility and digital caliper’s time response in skinfold measurement in male and female young adults | es |
| dc.type | journal article | es |
| dc.rights.accessRights | open access | es |
| dc.journal.title | Journal of the International Society of Sports Nutrition | es |
| dc.volume.number | 20 | es |
| dc.issue.number | 1 | es |
| dc.description.discipline | Actividad Física y Deporte | es |
| dc.identifier.doi | 10.1080/15502783.2023.2265888 | es |